AirSep Corporation • 401 Creekside Drive • Buffalo, NY 14228-2085 USATelephone: 716. 691.0202 • 24-Hour Fax: 716.691.4141 • www.airsep.comLifeStyle™

Seite 2 - In the US or Canada

1.6 LifeStyle ClassificationsType of protection against electric Protection from electric shock is achieved byshock: double insulation.Class IIDegree

Seite 3 - Table of Contents

2-12.0 Operation Check and Oxygen Concentration Test2.1 Description of OperationAir enters the LifeStyle Oxygen Concentrator through an external air i

Seite 4 - Section 6.0

2-22.3 Alarm SystemAn audible alarm sounds if LifeStyle has a low battery, senses no inspiration, or ifperformance outside of specifications is detect

Seite 5 - 1.0 Introduction

2-32.3.5 Battery Fully ChargedWhen LifeStyle is connected to an AC or DC outlet and is in the OFF setting, theyellow BATTERY light illuminates and dis

Seite 6

2.3.9 General MalfunctionIf a general malfunction of the LifeStyle unit occurs, a half-second alarm soundsevery half second, and the red ALARM light

Seite 7 - Symbol Description

3.0 Patient Instructions3.1 General InstructionsIt is important that patients thoroughly understand how to operate AirSep’s LifeStyleunit. This enabl

Seite 8

3-22. Remove the filter from the unit and wash it in a solution of warm waterand soap.3. Rinse the filter thoroughly, and remove excess water with a s

Seite 9 - 1.5 LifeStyle Specifications

4-14.0 Equipment Provider Maintenance4.1 Routine MaintenanceThe LifeStyle unit includes two filters and a compressor that require scheduled mainte-nan

Seite 10 - 1.6 LifeStyle Classifications

4-24.1.3 Compressor ReplacementAs the Equipment Provider, you are responsible for monitoring the life of thecompressor, which must be replaced after 3

Seite 11 - 2.2 Operation Check

4-3A list of undesirable chemical agents includes, but is not limited to, the following,according to the plastic manufacturer: alcohol and alcohol-ba

Seite 12 - 2.3 Alarm System

AirSep® is a registered trademark of AirSep Corporation.LifeStyle™ is a trademark of AirSep Corporation.MN110-1 Rev A 07/07Expedited Product Warr

Seite 14 - 2.3.9 General Malfunction

5-15.0 Service5.1 ComponentsThe design of AirSep’s Lifestyle Portable Oxygen Concentrator allows for easy accessand removal of most components. This a

Seite 15 - 3.0 Patient Instructions

5-23. Use a Phillips-head screwdriver to remove the two screws on thecenter housing near the handle and the screw in the hole in thecenter housing nea

Seite 16 - 3.2.2 Checking the Hour Meter

5.3.2 Compressor Assembly InstallationTo install the new compressor assembly, perform the compressor removalprocedure in reverse order.NOTEAll washers

Seite 18 - 4.2 Infection Control

6-16.0 Troubleshooting6.1 General TroubleshootingBefore reviewing the Troubleshooting Chart, the following steps may be useful to isolateany malfuncti

Seite 19

6.2 Troubleshooting ChartProblem Possible Cause SolutionUnit does not turn on, Wall switch power is off. Turn on wall switch power.and AC adapter ispr

Seite 20

Appendix Exploded DrawingsFigure 1: LifeStyle — Control Panel A-1Figure 2: LifeStyle — Internal, Top View A-2Figure 3: LifeStyle — Internal, Bott

Seite 21 - 5.0 Service

Ref No. Part No. Description6 FI109-1 Filter, Foam, 15PPI, 2.35 x 2.35 x 1/2”7 FI110-1 Filter, Felt, #10, Intake8 MI187-1 Misc Knob, Selector Switch9

Seite 22 - 5.3 Compressor Replacement

Ref No. Part No. Description5 HX011-2 Heat Exchanger6 MO016-2 Mount, Sandwich, Male-Female7 SW002-1 Switch, Temperature, 145F NC8 VA319-1 Valve, Assy,

Seite 23

AirSep Corporation i – table of contentsTable of ContentsSection 1.0Introduction1.1 Equipment Provider Responsibility 1-11.2 Important Notice and Symb

Seite 24

Figure A.3: LifeStyle — Internal, Bottom ViewA-3

Seite 25 - 6.0 Troubleshooting

ii – table of contents AirSep CorporationSection 4.0Equipment ProviderMaintenance4.1 Routine Maintenance 4-14.1.1 Air Intake / Gross Particle Filter 4

Seite 26 - 6.2 Troubleshooting Chart

1.0 Introduction1.1 Equipment Provider ResponsibilityAll Equipment Providers of the LifeStyle™ Portable Oxygen Concentrator must assumeresponsibilitie

Seite 27 - Appendix

1-2 Instruct patients to notify their physicians and/or Equipment Providers if theyexperience any signs of discomfort. Instruct each patient how to

Seite 28

1-31.3 Definitions of SymbolsThe definitions of symbols used with the LifeStyle Portable Oxygen Concentrator follow. Symbol DescriptionRefer to Patien

Seite 29

1-41.4 LifeStyle Product Limited WarrantyAirSep Corporation warrants the LifeStyle Oxygen Concentrator to be free fromdefect in parts for two years (o

Seite 30

1-51.5 LifeStyle SpecificationsOxygen Concentration:* Set at 1-5 LPM: equivalent to a continuous flow of 90%oxygen ± 3%Dimensions: 5.5 in. high x 7.2